Index: ssts-basedata/src/main/java/com/forgon/disinfectsystem/entity/sterilizationmanager/sterilizerrecord/XinhuaCleanH101513Details.java =================================================================== diff -u -r40941 -r40944 --- ssts-basedata/src/main/java/com/forgon/disinfectsystem/entity/sterilizationmanager/sterilizerrecord/XinhuaCleanH101513Details.java (.../XinhuaCleanH101513Details.java) (revision 40941) +++ ssts-basedata/src/main/java/com/forgon/disinfectsystem/entity/sterilizationmanager/sterilizerrecord/XinhuaCleanH101513Details.java (.../XinhuaCleanH101513Details.java) (revision 40944) @@ -25,12 +25,58 @@ public void setId(Long id) { this.id = id; } + /** + * 时间 - 数据记录的具体时间点(时分秒格式) + * 格式: "HH:mm:ss",示例: "13:16:38" + * 用于标识每个数据点的精确时间位置 + */ private String time; + + /** + * 日期时间 - 数据记录的完整时间戳 + * 格式: Date对象,包含年月日时分秒 + * 示例: 2025-07-15 13:16:38 + * 用于精确的时间追溯和数据排序 + */ private Date dateTime; + + /** + * 阶段 - 灭菌程序当前运行的阶段名称 + * 示例: "准备"、"脉动"、"跨压脉动"、"正压脉动"、"升温"、"灭菌"、"排汽"、"真空干燥"、"平衡"、"结束" + * 反映灭菌设备当前的工作状态和流程进度 + */ private String stage; + + /** + * 内室压力 - 灭菌室内腔的实际压力值(kPa) + * 单位: 千帕(kPa),示例: 0.4, -89.7, 212.5 + * 正值表示正压状态,负值表示真空状态 + * 关键参数:影响灭菌效果和安全性 + */ private double innerPressure; + + /** + * 内室温度 - 灭菌室内腔的实际温度值(℃) + * 单位: 摄氏度(℃),示例: 88.4, 134.3, 135.4 + * 核心参数:直接决定灭菌效果的关键指标 + * 必须达到并维持设定温度才能有效灭菌 + */ private double innerTemperature; + + /** + * 夹层压力 - 灭菌设备夹套层的压力值(kPa) + * 单位: 千帕(kPa),示例: 214.6, 221.0, 135.8 + * 夹层是灭菌室外的加热介质层 + * 用于控制和维持内室温度的压力系统 + */ private double interlayerPressure; + + /** + * 夹层温度 - 灭菌设备夹套层的温度值(℃) + * 单位: 摄氏度(℃),示例: 134.8, 136.2, 135.8 + * 加热介质的温度,通过夹层间接加热内室 + * 影响内室温度的加热效率和稳定性 + */ private double interlayerTemperature; private Long xinhuaCleanH101513Id;